Cool Drawer Single and Dual Sensors Operating Manual

1.5.2 Installation Requirements

Feedthrough. . . . . . . . . . . . . . . . . . . INFICON provided CF40 feedthrough or

2.54 cm (1 in.) bolt feedthrough

Other . . . . . . . . . . . . . . . . . . . . . . . . XIU or oscillator to match specific

controller/monitor
CDS-X1XXX only: Solenoid Valve for air
PN 750-420-G1 (see

section 3.1 on page

3-1

)

Water Flow Rate. . . . . . . . . . . . . . . . Minimum 150 to 200 cm

3

/min,

30°C maximum

Water Quality . . . . . . . . . . . . . . . . . . Coolant should not contain chlorides as

stress corrosion cracking may occur.
Extremely dirty water may result in loss of
cooling capacity.

CAUTION

Do not allow water tubes to freeze. This may happen if the
tubes pass through a cryogenic shroud and the flow of
fluid is interrupted.

Air (shutttered sensors only) . . . . . . 55 PSIG {70 PSIA} (3.8 bar)

[379 kPa](minimum)

60 PSIG {75 PSIA} (4.1 bar)
[414 kPa](maximum)

WARNING

Do not exceed 100 PSIG (115 PSIA) (6.9 bar) [689 kPa].
Connection to excessive pressure may result in personal
injury or equipment damage.