11

Fieldbus name

Profibus

Standard applied

→ [7]

Communication protocol data format

→ [9], [7]

Interface, physical

RS485

Data rate

≤12 MBaud (

→ [2])

Node address

Local
(Adjustable via hexadecimal
"ADDRESS", "MSD", "LSD"
switches)

Default

setting

Via Profibus
(hexadecimal "ADDRESS" switches
set to >7d

hex

(>125

dec

))





00 … 7D

hex

(0 … 125

dec

)

5C

hex



00 … 7D

hex

(0 … 125

dec

)

Profibus connection

D-Sub, 9-pin, female

Cable

Shielded, special Profibus cable
(

→ 22 and [5])

Cable length, system wiring

According to Profibus specifications
(

→ [7], [5])


Materials exposed to vacuum

Housing, supports, screens
Feedthroughs
Insulator
Cathode
Cathode holder
Pirani element
Sensor diaphragm
Sensor contacts


stainless steel
NiFe, nickel plated
glass
iridium, yttrium oxide (Y

2

O

3

)

molybdenum
tungsten, copper
ceramic (Al

2

O

3

)

SnAg

Internal volume

DN 25 ISO-KF
DN 40 CF-R


≈24 cm

3

≈34 cm

3

Pressure max.

5 bar (absolute)


Admissible temperatures

Storage
Operation
Bakeout

-20 … 70 °C
0 … 50 °C
+150 °C (at vacuum connection, without
electronics unit, horizontally mounted

Relative humidity
(year's mean / during 60 days)


≤65 / 85% (no condensation)

Use indoors

only

altitude up to 2000 m NN

Mounting orientation

any

Degree of protection

IP 30


Profibus interface
(BCG450-SP)

Materials used