INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual

Triplegauge, Safety, Bcg450-sd

Advertising
background image

TripleGauge™

Bayard-Alpert Pirani Capacitance Diaphragm Gauge
With Fieldbus Interface

BCG450-SD
BCG450-SP

Instruction Sheet

ti

ma41e1-b (2011-04)



About this document

This document is a supplement to the standard Instruction
Sheet enclosed with the BCG450 ( [1]). It should be
used together with the standard Instruction Sheet.
The symbol (

→ [XY]) refers to documents and files

listed under "Further Information".


Product Identification

In all communications with INFICON, please specify the in-
formation on the product nameplate. For convenient refer-
ence copy that information into the space provided below.

Model:
PN:
SN:
V W

INFICON AG, LI-9496 Balzers


Validity

This document applies to products with the following part
numbers:

BCG450-SD (DeviceNet):

353-557

(DN 25 ISO-KF)

353-558

(DN 40 CF-R)

353-562

(DN 25 ISO-KF, with baffle)

BCG450-SP (Profibus):

353-554

(DN 25 ISO-KF)

353-556

(DN 40 CF-R)

The part number (PN) can be taken from the product name-
plate.
If not indicated otherwise in the legends, the illustrations in
this document correspond to the vacuum connection
DN 25 ISO-KF. They apply to other vacuum connections by
analogy.
We reserve the right to make technical changes without prior
notice.
All dimensions in mm.

Trademarks

DeviceNet™

Open DeviceNet Vendor Association, Inc.

TripleGauge™ INFICON

AG,

Balzers


Intended Use

The BCG450-SD and BCG450-SP gauges have been de-
signed for vacuum measurement of gases in the pressure
range 5×10

-10

… 1500 mbar.

They must not be used for measuring flammable or combus-
tible gases in mixtures containing oxidants (e.g. atmospheric
oxygen) within the explosion range.




Safety

Symbols Used

DANGER

Information on preventing any kind of physical injury.

Caution

Information on correct handling or use. Disregard can lead
to malfunctions or minor equipment damage.


Personnel Qualifications

Skilled personnel

All work described in this document may only be carried
out by persons who have suitable technical training and
the necessary experience or who have been instructed by
the end-user of the product.


General Safety Instructions

All safety instructions given in [1] and [2] apply to the
sensor types described in this document, too.

Liability and Warranty

INFICON assumes no liability and the warranty becomes null
and void if the end-user or third parties
• disregard the information in this document
• use the product in a non-conforming manner
• make any kind of changes (modifications, alterations etc.)

to the product

• use the product with accessories not listed in the product

documentation.

The end-user assumes the responsibility in conjunction with
the process media used.

Gauge failures due to contamination or wear and tear, as
well as expendable parts (e.g. filament), are not covered by
the warranty.

BCG450-SD

General Information

The BCG450-SD gauge has a fieldbus interface that
conforms to the DeviceNet standard (

→ [8]).

Via this interface, the following and further data are
exchanged in the standardized DeviceNet Protocol
(

→ [3], [8]):

• Pressure reading
• Pressure unit (mbar, Torr, Pa)
• Degas function
• Status and error messages

Two adjustable switching functions are integrated in the
gauge. With the built-in atmosphere switching function an
atmospheric pressure thresold can be programmed
(

→ [2]). The corresponding relay contacts are available at

the sensor cable connector.

The basic sensor and sensor electronics of the BCG450-SD
type are the same as in the standard BCG450 (

→ [1], [2]).



Technical Data

General technical data of the sensor and sensor

electronics

→ [1], [2]

Fieldbus Interface

Fieldbus name

DeviceNet

Standard applied

→ [8]

Communication protocol,
data format


→ [3], [8]

Interface, physical

CAN bus

DeviceNet Parameters

Data rate (adjustable via
"RATE" switch)

125 kBaud
250 kBaud
500 kBaud (default)
"P" (programmable
125 kBaud, 250 kBaud,
500 kBaud via DeviceNet
(

→ [3])

Node address (MAC ID)
(adjustable via "ADDRESS,
MSD, LSD" switches)

0 … 63

dec

(default 63

dec

)

"P" (programmable
0 … 63

dec

via DeviceNet)

(

→ [3])

DeviceNet connector

Micro-Style, 5-pin, male

Cable Shielded

special

DeviceNet cable,
5 conductors
(

→ [6], [8])

Cable length, system wiring

According to DeviceNet
specifications
(

→ [6], [8])

Supply Voltages

Supply voltage at the sensor
connector, Pin 8


+24 VDC (+20 … +28 V)

Power consumption

<20 W

DeviceNet operation requires an additional,

separate power supply.

DANGER

The gauge may only be connected to power
supplies, instruments or control devices that
conform to the requirements of a grounded pro-
tective extra-low voltage (SELV). The connection
to the gauge has to be fused.

Supply voltage at the DeviceNet
connector, Pin 2


+24 VDC (+11 … +25 V)

Power consumption

<2 W

The gauge is protected from reversed polarity of the supply
voltage.

Advertising