Adjustment and Settings

For BCG450-SD and BCG450-SP gauges.

Adjusting the Gauge

The gauge is factory calibrated. If used under different cli-
matic conditions, at extreme temperatures, through aging or
contamination and after exchanging the sensor, the charac-
teristic curve can be offset and readjustment can become
necessary. However, this adjustment is carried out automati-
cally during operation by the gauge itself. No manual adjust-
ment is required (adjustment of the atmosphere sensor
→ [2]).

Setting the Switching Functions

The threshold values of switching functions A and B

1)

can be

set within the pressure range 1×10

-9

mbar … 100 mbar via

potentiometers "SETPOINT A" and "SETPOINT B". For the
corresponding threshold voltages U

Threshold

, the following

equation applies:

U

Threshold

= 0.75 × (log p

Setpoint

– c) + 7.75

Constant c depends on the pressure unit (

→ [1], [2]).

1)

Relays SP A/B can be reprogrammed for atmosphere

switching function via fieldbus interface (

→ [2]) (default

switching function A/B).

Time t

Mea

sured

value

Off

Off

On

Hysteresis
10% U

Threshold

Switching function

U

Threshold

Measuring signal
(Pressure p)

(Setpoint A, B)

The hysteresis of the switching function is 10% of the thresh-
old setting.

n

Put the gauge into operation.

o

Connect the + lead of a voltmeter to the threshold

measurement point of the selected switching function
("Setpoint A" Pin 3, "Setpoint B" Pin 6) and its lead to
a ground contact nearby (eg. grounded locking screw
nut of connector or vacuum connection of the gauge).

The threshold voltages are referenced to

ground (housing, vacuum connection), not to
Pin 5 (common power GND 24 V supply).

max. ø2.5

Setpoint A Pin 3
Setpoint B Pin 6

( )

p

Using a screwdriver (max. ø2.5 mm), set the threshold

of the selected switching function (Setpoint A, B) to the
desired value U

Threshold

.

A functional check of the switching functions (On/Off) is only
possible via fieldbus interface (

→ [3] for BCG450-SD,

→ [4] for BCG450-SP) or by measuring the relay contacts

with a continuity checker/ohmmeter (

→ "Electrical Connec-

tion", sensor cable connector).

Further Information

[1] www.inficon.com

Instruction sheet
TripleGauge™ BCG450
tima40d1 German
tima40e1 English
INFICON AG, LI–9496 Balzers, Liechtenstein

[2] www.inficon.com

Instruction manual
TripleGauge™ BCG450, BCG450-SD, BCG450-SP
tina40d1 German
tina40e1 English
INFICON AG, LI–9496 Balzers, Liechtenstein

[3] www.inficon.com

Communication protocol
DeviceNet™ BCG450-SD

tira40e1 English
INFICON AG, LI–9496 Balzers, Liechtenstein

[4] www.inficon.com

Communication protocol
Profibus BCG450-SP

tira41d1 German
tira41e1 English

INFICON AG, LI–9496 Balzers, Liechtenstein

[5] www.inficon.com

("Semiconductor and Vacuum coating processes,
Vacuum Gauges")
Product descriptions and downloads
INFICON AG, LI–9496 Balzers, Liechtenstein

[6] www.odva.org

Open DeviceNet Vendor Association, Inc.
"DeviceNet™ Specifications"

[7] www.profibus.com

Profibus user organisation

[8] European Standard for DeviceNet EN 50325

[9] European Standard for Profibus EN 50170



LI–9496 Balzers
Liechtenstein
Tel +423

/

388

3111

Fax +423 / 388 3700
[email protected]

www.inficon.com

Advertising
This manual is related to the following products: