Installation, Measuring signal vs. pressure, Gas type dependence – INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 2: Vacuum connection, Electrical connection

Advertising
background image

Voltage at gauge

+24 VDC (+20 … +28 VDC)
(ripple

≤2 V

pp

)

2)

Power consumption

Standard
Degas
Emissions start (200 ms)


≤0.5 A

≤0.8 A

≤1.4 A

Fuse required

1)

≤1.25 AT

Power consumption

≤18 W (BPG402-S/-SL only)

Electrical connection

D-Sub, 15 pins, male

Sensor cable

Shielded, number of
conductors depend on
functions used.

Cable length (24 VDC)

≤35 m (0.25 mm²/conductor)

≤50 m (0.34 mm²/conductor)

≤100 m (1.0 mm²/conductor)

For operation with
RS232C interface


≤30 m

Materials on the vacuum side

Housing, supports,
screens
Feedthroughs
Insulator
Cathode
Cathode holder
Pirani element



stainless steel
NiFe nickel plated
glass
iridium, yttrium oxide (Y

2

O

3

)

molybdenum, platinum
tungsten, copper

Internal volume

DN 25 ISO-KF
DN 40 CF-R


≈24 cm

3

≈34 cm

3

Pressure max.

≤2 bar (absolute)

Admissible temperatures

Storage

Operation

Bakeout
Long

tube

–20 … +70 °C

0 … +50 °C

80 °C

3)

150 °C

3)

Relative humidity

Year’s mean
During 60 days


≤65% (not condensable)

≤85% (not condensable)

Use indoors

only

altitude up to 2000 m NN

Mounting orientation

any

Type of protection

IP 30

1)

INFICON controllers fulfill these requirements.

2)

Consider the voltage drop on the sensor cable.

3)

Flange temperature, without electronics unit, horizontally
mounted.

Dimensions

DN 25 ISO-KF

DN 40 CF-R

DN 40 CF-R

long tube

15

5

.2

11

1

59

67

48

42

Weight 353-570,

353-572

353-571, 353-573
353-578

450 g

710 g

917 g


Measuring Signal vs. Pressure

1E+01

1E–01

1E–03

1E–05

1E–07

1E–09

1E+03

0.0

se

nso

r e

rr

o

r

ove

rra

nge

1E+04

1E+02

1E+00

1E–02

1E–04

1E–06

1E–08

1E–10

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

10.0

un

der

rang

e

Measuring signal U [V]

Pressure p [mbar]

p = 10

(U-7.75)/0.75+c

U p c

[V] [mbar] 0

[V] [Pa] 2

[V] [Torr]

-0.125

where p

pressure

U measuring

signal

c

constant (pressure unit dependent)

Gas Type Dependence

For gases other than air, the pressure in the indication range
p < 10

-3

mbar can be determined by a simple conversion:

p

eff

= C × pressure indicated

Gas
type

Calibration

factor C

Gas

type Calibration

factor C

He
Ne

Kr
Ar

5.9
4.1
0.5
0.8

air,

O

2

, CO, N

2

H

2

Xe

1.0
2.4
0.4




Installation

Vacuum Connection

DANGER

DANGER: overpressure in the vacuum system
>1 bar

Injury caused by released parts and harm
caused by escaping process gases can result if
clamps are opened while the vacuum system is
pressurized.

Do not open any clamps while the vacuum sys-
tem is pressurized. Use the type clamps which
are suited to overpressure.

DANGER

DANGER: protective ground

Incorrectly grounded products can be extremely
hazardous in the event of a fault.

The gauge must be electrically connected to the
grounded vacuum chamber. This connection
must conform to the requirements of a protective
connection according to EN 61010:
• CF connection fulfill this requirement
• For gauges with a KF flange, use a conduc-

tive metallic clamping ring

Caution

Caution: vacuum component

Dirt and damages impair the function of the
vacuum component.

When handling vacuum components, take ap-
propriate measures to ensure cleanliness and
prevent damages.

Caution

Caution: dirt sensitive area

Touching the product or parts thereof with bare
hands increases the desorption rate.

Always wear clean, lint-free gloves and use
clean tools when working in this area.

The gauge may be mounted in any orientation. To

keep condensates and particles from getting into the
measuring chamber, preferably choose a horizontal
to upright position.

The gauge is supplied with a built-in grid. For poten-
tially contaminating applications and to protect the
electrodes against light and fast particles, installation
of the optional baffle is recommended (

→ [1]).

Vacuum connection free of grease.

Remove the protective lid and install the gauge to the
vacuum system. Keep the protective lid.

Electrical Connection

(BPG402-S/-SL only)

Make sure the vacuum connection is properly made

(

→ "Vacuum Connection").

If no connection cable is available, make one according

to the following diagram.

Common (power GND 24V supply)

Ground (housing, vacuum connection)

8

9

1

15

D-Sub,15 pins,

female,

soldering side

SP

TxD

RxD

Degas

42 k

Ω

3

4

1

11

9

13

14

7

8

2

12

5

15

1.25 AT

24V

Degas

Ident.

RS232

10

-

-

-

1)

( )

Measuring

signal

-

V

S

Threshold value, SP

Filament status

6

Electrical connection

Pin 1 Relay switching function,

common contact

Pin 2 Measuring signal output

0 … +10 V

Pin 3 Threshold (setpoint)

1)

0 … +10 V

Pin 4 Relay switching function,

n.o. contact

Pin 5 Supply common

0 V

Pin 6 Not connected internally

Pin 7 Degas (active high)

0 V/+24 V

Pin 8 Supply (V

s

) +24

V

Pin 9 Relay filament status

common contact

2)

Pin 10 Gauge identification

Pin 11 Relay filament status,

n.o. contact

2)

Pin 12 Measuring signal common

Pin 13 RS232C, TxD

Pin 14 RS232C, RxD

Pin 15 Do not connect

1)

Do not connect pin 3 for normal operation of the

gauge. This pin is reserved for adjustment of the
setpoint potentiometer (

→ section "Switching

Function").

2)

→ section "Filament Status"


t i ma 4 6 e 1 - b

(2010-03)

Original: German tima46d1-b (2010-03)