Protocol description, Communication error, Output pressure error – INFICON PSG55x ATM to Medium Vacuum Gauge (RS232C / RS485C) User Manual

Page 5

Advertising
background image

tira59e1 (2010-07) RS232C.cp

5

If a communication error occurs during transmission the PID is set to 0xFFFF and
an error byte is added.

Byte 0, errors

Error Code Description

1 Access

error

2

Value higher max. or lower min.

3

Parameter not found

4 Length

error

6 Memory

access

error

7 Memory

access

timeout


The following parameters can be read via a serial interface.

PID Name

Description

Factory

setting

Min.

Max.

Type

221 Pressure

Output pressure in integer format

Fixs32en20

222 Pressure

Pressure in vacuum chamber in Real
format

Real32

265 ATM Pressure

Ambient pressure in real format

Real32

466 Differential Pressure

Differential pressure outside and inside
vacuum chamber

Real32

224 Data Unit

Changes the pressure unit of all Real32
pressures.

0 mbar

1 Torr

2 Pascal

3 micron

4 Counts

0

0

4

Uint8

PID Name

Description

Factory

setting

Min.

Max. Type

228 Device Exception

0 No

error

1

Timeout EEPROM memory access

2

EEPROM CRC error

3 EEPROM

error

4

Pirani filament rupture

5 Wrong

filament

material

6 CDG

diaphragm

rupture

8

ATM outside specification limits

11 Sensor does not match gauge

0

Uint8