General information sensor details – INFICON PSG55x ATM to Medium Vacuum Gauge (RS232C / RS485C) User Manual

Page 6

Advertising
background image

6

tira59e1 (2010-07) RS232C.cp

PID Name

Description

Factory

setting

Min.

Max.

Type

103 Reset

0 Write

The sensor carries out a reset.

1 Write

The sensor resets all parameters to their
factory settings.

104 Run Hours

Operating hours [0.25 h]

Fixs32en2

207 Serial Number

Serial number

4294967295 Uint32

208 Product Name

Product name

PCG550

String

209 Manufacturers Name

Name of manufacturer

INFICON AG

String

210

Manufacturers Model
Number

Model number

String

218 Software Version

Software version

String

227 RS232 Baud Rate

9600, 19200, 38400, 57600

57600

9600

57600

Uint32

243 Display Direction

Display orientation

0

Setting for flange in bottom position

1

Setting for flange in top position

0

0

1

Uint8



PID

Name

Description

Factory

setting

Min.

Max. Type

223

Active Instance Number

Current sensor

1

CDG sensor active

2

Pirani sensor active

3

Mixed signal range Pirani and CDG
sensor active

Uint8

33000 Pirani Full Scale

Pirani fullscale

1000

Fixs32en20

33001 Pirani Overrange Value

Pirani overrange status output limit

1000

Fixs32en20

33002 Pirani Underrange Value

Pirani underrange status output limit

5.00E-05

Fixs32en20

255

Pirani Safe State

Pressure reading in mbar that is output in
the event of an error in Pirani mode

0 0

mbar

1 1500

mbar

2

The last valid value is retained.

3

The Pirani Safe State value is output.

0

0

3

Uint8

256

Pirani Safe State Value

Pressure reading in mbar that is output in
the event of an error if the Pirani Safe State
is set to 3.

0

0

2047 Fixs32en20

417

Pirani Adjust Flag

Executes a manual Pirani sensor adjust-
ment.
Set the value to "1" in order for the adjust-
ment to be executed.

0

Uint8

236

CDG Safe State

*

Pressure reading in mbar that is output in
the event of an error in CDG mode.

0 0

mbar

1 1500

mbar

2

The last valid value is retained.

3

The CDG Safe State value is output.

0

0

3

Uint8

237

CDG Safe State Value

*

Pressure reading in mbar that is output in
the event of an error if the CDG Safe State
is set to 3.

0

0

2047 Fixs32en20