Assembly objects (class code 04 hex ), Output assemblies, Input assemblies – INFICON PEG100-D Medium to Ultra-High Vacuum Gauge User Manual

Page 6: 3 assembly objects (class code 04

Advertising
background image

6

tina12e1 (0106) PEG100-D.ga

Attribut ID

Access Rule

Name

Data/Type

Description

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
55 (37

hex

)

get

Serial Number

String [5]

56 (38

hex

)

get

Device Configuration

String [8]

PEG100-D

57 (38

hex

)

get

Device Status

UNIT

Device status
1 = Initialising
2 = Idle (HV on)
4 = Executing (HV on)

58 (3A

hex

)

get / set

Reporting Mode

BYTE

Polling, Bit Strobe = 6
COS / Cyclic = 0

60 (3C

hex

)

get

Exception Status

BYTE

0

hex

= ok

1

hex

= HV on, no plasma

2

hex

= HV off

Services

Service Code

Name

Description

–––––––––––––––––––––––––––––––––––––––––––––––––
14 (0E

hex

)

Get Attribute Single

3.3 Assembly Objects (Class Code 04

hex

)

A collection of assembly objects allows the sending of attributes from different application objects in one message
(i.e.: Polling I/O).

3.3.1 Output Assemblies

Messages which a master sends to the PEG100-D.

Output Assembly 1

Byte

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
0

res

res

res

res

res

res

HV

HV

ON / OFF

ON / OFF
Control

3.3.2 Input Assemblies

Messages which the PEG100-D sends to the master.

Input Assembly 2

Byte

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
0

res

res

res

res

res

HV

HV Status

Sensor

ON / OFF

Satus

Source

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
1

Exception Status

2

Pressure value (Low Byte)

3

Pressure value (Low Middle Byte)

4

Pressure value (High Middle Byte)

5

Pressure value (High Byte)