2.2.1 Meaning of PEG100-P (slave) input and

output data

2.2.1.1 PEG100-P output data

(output data from the master’s view point)

The following two 2 bytes are sent to the device:

Output data

———————————————–

Byte 0

Bit 7

not used

Bit 6

not used

Bit 5

not used

Bit 4

not used

Bit 3

not used

Bit 2

not used

Bit 1

Control Bit

Bit 0

Control Bit

———————————————–

Byte 1

not used

Semantics

Byte 0 / Bit 0 and Bit 1

The high voltage could be switched on by the digital
inputs (Pin 7 or Pin 8) at the FCC 68 plug or directly by
a profibus command. The input has to be selected by
these two bits.

Bit 1

Bit 0 Meaning

HEX-
Value

————————————————————————–

0

0

Digital Inputs active

0

1

0

Fieldbus active, High Voltage OFF

2

1

1

Fieldbus active, High Voltage ON

3

2.2.1.2 PEG100-P input data

(from master’s view point)

The input data contain the following information: status
information and measurement results (as 12-bit values).

Input data

—————————————————————–

Byte 0

Bit 7

Status bit

Bit 6

Status bit

Bit 5

Status bit

Bit 4

Status bit

Bit 3

Measurement value: Bit 11: MSB

Bit 2

Measurement value Bit 10

Bit 1

Measurement value Bit 9

Bit 0

Measurement value Bit 8

———————————————–——————

Byte 1

Bit 7

Measurement value Bit 7

Bit 6

Measurement value Bit 6

Bit 5

Measurement value Bit 5

Bit 4

Measurement value Bit 4

Bit 3

Measurement value Bit 3

Bit 2

Measurement value Bit 2

Bit 1

Measurement value Bit 1

Bit 0

Measurement value Bit 0: LSB

Explanation

Byte 0 / Bit 4 - Bit 7 - Status messages

Bit 7 Bit 6 Bit 5 Bit 4 meaning

HEX-
Value

———————————————————————————

1

0

1

0

High Voltage OFF

A

1

1

1

0

High Voltage ON,

E

No Ignition

1

1

0

0

High Voltage ON,

C

Measurement Value ok

0

1

1

0

HV on, but no ignition

6

after 1 min of HV on

Measurement values - representation
The measurement value is given as a 12-bit value. Byte
0/Bit 3 represents the MSB, byte 1/Bit 0 the LSB.

How to calculate the measurement value
The measurement values are calculated according to the
following formulas in the corresponding measurement
units.

Calculation formula in unit „mbar“

pressure p / mbar= 10

(output value / 500 - 9.5)

A special pressure’s value is calculated with the follo-
wing formula:

value = 500 · (log (p / mbar) + 9,5)

Calculation formula in unit „torr“

pressure p = 10

(output value / 500 - 9.6249)

A special pressure’s value is calculated with the follo-
wing formula:

value = 500 · (log (p / Torr) + 9,6249)

Calculation formula in unit „Pascal“

pressure p = 10

(output value / 500 - 7.5)

A special pressure’s value is calculated with the follo-
wing formula:

value = 500 · (log (p / Pa) +7,5)

6