Installation, Measuring signal vs. pressure, Gas type dependence – INFICON MPG40x ATM to Ultra-High Vacuum Gauge User Manual

Page 2: Vacuum connection, P = 10, U = c + 0.6log, K × pressure reading

Advertising
background image

Internal volume

20 cm³

Pressure

10 bar (absolute),

limited to inert gases

Temperatures

Operation

2)

Bakeout

Pirani filament
Storage


+5…+55 °C
150 °C (without electronics
and magnetic shielding)
120

°C

-40…+65 °C

Relative humidity

≤80% at temperatures

≤+31°C decreasing to 50%

at +40°C

Use indoors

only

altitude up to 2000 m

Type of protection

IP 40

Dimensions [mm]

18

DN 25 ISO-KF

18

.5

DN 40 ISO-KF

DN 40 CF-F

24

92

ш 63.5

28

55

60

75

10

5

32

55

60

75

15

.5

ш 63.5


Weight

351-010
351-011
351-012


≈700 g

≈720 g

≈980 g


351-020
351-021
351-022


≈730 g

≈750 g

≈1010 g

Measuring Signal vs. Pressure

Pressure p

1E+04

1E+02

1E+00

U [V]

0.00.5 1.5 2.5 3.5 4.5 5.5 6.5 7.5 8.5 9.5 10.5

1E–02

1E–04

1E–06

1E–08

sen

sor

er

ror

un

der

ran

ge

ove

rra

nge

sen

sor

er

ror

Pa
mbar

torr

p = 10

1.667U-d

Measurement signal U [V]

p = 10

1.667×U-d

U = c + 0.6log

10

p

mbar Pa Torr

d 11.33 9.33 11.46
c 6.8 5.6 6.875

valid in the range 5×10

-9

mbar <p< 1000 mbar

3.8×10

-9

Torr <p< 750 Torr

5×10

-7

Pa <p< 1×10

5

Pa


2)

MPG401: Up to 150 °C at the flange if mounted horizontally;
without magnetic shielding

Gas Type Dependence

10

2

8

6
4

2

10

1

8

6
4

2

10

0

8

6
4

2

10

–1

8

6
4

2

10

–2

p (mbar)

p

eff

(mbar)

10

–3

2

4 6

10

–2

2

4 6

10

–1

2

4 6

10

0

2

4 6

10

1

2

4 6

10

2

H

2

He Ne

Xe

Ar

Freon 12

Water
vapor

Air
O2

CO

N2

Kr

Indication range
>10

-2

mbar)

CO

2

p (mbar)

6
4

2

6
4

2

6
4

2

10

–4

8

6
4

2

He

Ar

Air

N2

O2

CO

10

–5

8

6
4

2

10

–6

10

–3

8

10

–2

8

10

–1

8

Indication range
10

-6

... 10

-1

mbar

p

eff

(mbar)

10

-6

2

4 6

10

-6

2

4 6

10

–4

2

4 6

10

-3

2

4 6

10

-2

2

4 6

10

-1

In the range below 10

-5

mbar, the pressure indication is

linear.
For gases other than air, the pressure can be determined by
means of a simple conversion formula:

p

eff

= K × pressure reading

Gas type

Air

(O

2

, CO, N

2

)

Xe Kr Ar H

2

Ne He

K (mean
values)

1.0 0.4 0.5 0.8 2.4 4.1 5.9





Installation

Vacuum Connection

DANGER

DANGER: overpressure in the vacuum system
>1 bar
Injury caused by released parts and harm
caused by escaping process gases can result if
clamps are opened while the vacuum system is
pressurized.
Do not open any clamps while the vacuum sys-
tem is pressurized. Use the type of clamps which
are suited to overpressure.

DANGER

DANGER: overpressure in the vacuum system
>2.5 bar
KF flange connections with elastomer seals (e.g.
O-rings) cannot withstand such pressures. Pro-
cess media can thus leak and possibly damage
your health.
Use O-rings provided with an outer centering
ring.

DANGER

The gauge must be electrically connected to the
grounded vacuum chamber. This connection
must conform to the requirements of a protective
connection according to EN 61010:
• CF connections fulfill this requirement.
• For gauges with a KF flange, use a conduc-

tive metallic clamping ring

Caution

Caution: vacuum component
Dirt and damages impair the function of the
vacuum component.
When handling vacuum components, take ap-
propriate measures to ensure cleanliness and
prevent damages.

Caution

Caution: dirt sensitive area
Touching the product or parts thereof with one's
bare hands increases the desorption rate.
Always wear clean, lint-free gloves and use
clean tools when working in this area.

WARNING

WARNING: electric arcing
Helium may cause electric arcing with detrimen-
tal effects on the electronics of the product.
Before performing any tightness tests put the
product out of operation and remove the elec-
tronics unit.

The gauge may be mounted in any orientation. To

keep condensates and particles from getting into the
measuring chamber preferably choose a horizontal
to upright position and possibly use a seal with a
centering ring and filter.

When making a CF flange connection, it can be

advantageous to temporarily remove the electronics
and the magnet unit (

Operating Manual

tina48e1).

Remove the protective lid and install the product at the
vacuum system.

Hex. socket

AF 1.5

Protective lid

AF 7

Keep the protective lid.



t i ma 4 8 e 1 - a

(2008-10)

Original: German tima48d1-a (2008-10)

Advertising