Keep the button depressed

max. ø1.1 mm

Push the <ZERO> button again:

Fine adjustment
within 0...3 s:

the zero adjustment value changes
by one unit (push <ZERO> button in
intervals of 1 s)

Change of direction
within 3...5 s:

the zero adjustment changes its
direction (the blinking frequency of
the <ERROR> LED changes briefly)

If the <ZERO> button is released for more than 5 s, the

gauge returns to the measurement mode.

The <ERROR> LED will continue blinking red if the signal output
is negative (< -20 mV).