2 technical data – INFICON Cube CDGsci User Manual
Page 9

tina88e1-b (2014-09)
9
2 Technical
Data
Measurement range
→ 2, "Validity"
Accuracy
1)
0.1 F.S.
1 … 1000 F.S.
0.05% of reading
0.025% of reading
Temperature effect on zero
0.1 F.S.
1 … 1000 F.S.
10 ppm F.S./ °C
5 ppm F.S./ °C
Temperature effect on span
0.1 F.S.
1 … 1000 F.S.
10 ppm of reading / °C
5 ppm of reading / °C
Resolution, nominal
0.95 ppm F.S.
Gas type dependence
none
Output signal analog (measurement signal)
Measurement range
0 … +10 V
Voltage range
–2 … +10.24 V
(limited to +10.20 V)
Relationship voltage-pressure
linear
Output impedance
0
Ω (short-circuit proof)
Loaded impedance
>10 k
Ω
Response time
2)
350 ms
Pressure unit
Torr (default), mbar, Pa
(
→ 16)
1)
Non-linearity, hysteresis, repeatability in the calibrated range at 25 °C
ambient operating temperature without temperature effects after
operation of 2 h.
2)
Increase 10 … 90 % F.S.R.