5.7. Setting the temperatures
Setting the working temperature:
1.
Press the setpoint key
.
The indicator light blinks and the effective value appears
on the MULTI-DISPLAY (LED).
2.
Use the cursor keys
to move left or right on the
display until the numeral you wish to change is blinking.
3.
Use the increase/decrease arrows
to change the
selected numeral (-, 0, 1, 2, 3, ... 9).
4.
Press enter
to store the selected value (example: 37.0
°C).
The working temperature is maintained constant after a short
heat-up time (example: 37.0 °C).
5.8. Setting the PID control parameters Xp, Tn and Tv
Setting the Xp (proportional range)
1.
Press the key
.
The indicator light blinks and the effective Xp value
appears on the MULTI-DISPLAY (LED).
2.
Use the cursor keys
to move left or right on the
display until the numeral you wish to change is blinking.
3.
Use the increase/decrease arrows
to change
the selected numeral (-, 0, 1, 2, 3, ... 9).
4.
Press enter
to store the selected value (example: 12
°C).
Setting range for the Xp parameter: 0.1 to 50 °C