LC4
27
6.6. Interface parameters: r - br - P - H
RS232C
Whenever the parameter setting of the controller is not conform to
those of the connected personal computer, a modification is
necessary.
Factory setting of the RS232 interface:
Baudrate:
4800
Bauds
Parity:
2 (even)
Handshake: 1 (Hardware handshake)
Data bits:
7
Stop bits:
1
Press the cursor key
until the desired parameter appears on
the MULTI-DISPLAY (LED).
Use the edit keys
to set the desired parameter and
then press enter
.
Adjustable interface parameters
REMOTE
0 = keypad control mode
1 = remote control mode via RS232C
BAUDRATE 12 = 1200 bauds
24 = 2400 bauds
48 = 4800 bauds
96 = 9600 bauds
PARITY
0 = no parity
1
=
odd
2
=
even
HANDSHAKE
0 = Protocol Xon/Xoff
(software handshake)
1 = Protocol RTS/CTS
(hardware handshake)
Like all parametes which can be entered through the keypad,
interface parameters are stored in memory even after the controller
is turned off.