JULABO LC 4 Temperature controller User Manual
Page 22

Operating procedures
22
5.9. Setting the safety temperature (with shutdown function)
(excess temperature protection)
Press the key to indicate the safety temperature value on
the MULTI-DISPLAY and using a screwdriver
simultaneously turn the setting screw to the desired
value (example: 50 °C).
Setting range:
0 °C to 360 °C
in 2 °C steps
This safety installation is independent of the control circuit.
As soon as this safety installation is triggered, a complete
shutdown of the connected heating device is effected.
!
+
The alarm is indicated by optical and audible signals
(continuous tone)
and
on the MULTI-DISPLAY (LED) appears the error message
"Error 01".
Recommendation:
Set the safety temperature at 5 to 10 °C above the working
temperature setpoint.
The excess temperature protection should be set at least 25 °C below the fire
point of the bath liquid used!
In the event of wrong setting there is a fire hazard!
We disclaim all liability for damage caused by wrong settings!