JULABO TW20 Water Baths User Manual

Page 17

Advertising
background image

Water Baths

17

6.4.

Maintaining a constant water level / Counter cooling

For cooling tasks near the ambient air temperature the liquid

level/cooling set can be used for counter cooling.
By special pipe routing, cool faucet water is continuously supplied to

the water bath, while at the same time, the heated water is drained

via the overflow connection of the Level/Cooling set.

A specific water flow rate of 100 ml/minute is sufficient to

compensate for the characteristic temperature.

Caution:

Securely attach all tubing to prevent slipping.

Observe the laws and regulations of the water distribution company valid in

the location where the unit is operated.

Use of the liquid level/cooling set for a continuous supply of

faucet water:
1. to keep the water level constant, especially for

applications up to the boiling point (supply of faucet

water only in the amount of evaporation losses).

2. for counter cooling of cooling tasks near the ambient

surrounding temperature (cool faucet water is

continuously supplied to the water bath, while at the

same time, the heated water is drained via the overflow

connection of the liquid level/cooling set).

Liquid level/cooling set

Order number: 8 970 415

11 drainage screw on water bath
20 compensation reservoir
21 connecting sleeve
22 supply/drainage sleeve
23 overflow sleeve
24 adaptor screw for constant liquid level function
25 adaptor screw assy. for counter cooling function and

simultaneous constant liquid level control

26 adjuster screw for filling level adjustment