Kds ezflow 2030, Operation procedures – KD Scientific EZFLOW 2030 User Manual

Page 6

Advertising
background image

5

KDS EZFlow 2030

LOAD SYRINGE

Fill the syringe with desired medication. Load the syringe onto the pump. (Refer to Figure 2).

Place the syringe into the bed. Be sure to insert the nozzle into the syringe slot. Then press

and hold the secure button on the driver to bring the driver block into such a position that

the extremity of the syringe plunger can be placed into the driver slot. Fasten the rubber

strip to secure the syringe.

ALARMS AND INDICATOR LIGHTS

SET TIME - Activated only in STOP mode. When pressed, sets the desired time. `SET

TIME’ indicator should illuminate.

PURGE - Only functions in STOP mode. When pressed, starts purging tubing set.

RUN/STOP - Press to start or stop the pump. `MODE’ indicator should illuminate.

NEAR END - When there is approximately one minute of infusion time remaining,

this indicator flashes and the alarm buzzer sounds. To silence the alarm, press the

`RUN/STOP’ button.

COMPLETE/OCCLUSION - When infusion is complete, the indicator is illuminated in

red and the alarm buzzer sounds.

When the device senses an occlusion by not being able to drive with a force of

6.5kg ± 1 kg, the indicator is illuminated in red and the alarm buzzer sounds. To

silence the alarm, press the `RUN/STOP’ button.

The battery indicator light (ornage) indicates the driver is using built-in battery power.

this light and AC power light (green) will blink together if battery is low.

AC power light (green) indicates the pump is connected to mains AC power and is

charging the battery.

Operation Procedures:

(Continued)

KEYPAD DESCRIPTION