4.

Instrument Setup

36

Place connecting pieces (32) for the hoses on
the spouts at the correct angle, and seal off
the inlet you do not need using the plug (34)
provided.

The connecting pieces (32, 34) have to
be installed BEFORE the cover is at-
tached to the main filter and before the
prefilter is attached to the side of the
filter housing. Failure to observe this
instruction can result in breakage of
the cover and the prefilter housing.

Place the lid flat on a level surface and push
in the connecting pieces firmly.

Be absolutely sure not to kink the ex-
haust hose, as otherwise the suction
generated by the filter will be impaired.

32

32

32

34

The short black hose (36) connects the prefil-
ter (37) to the main filter in the filter unit.

The longer gray hose (38) connects the prefil-
ter to the laser printer.

Only now is the power cable (35) plugged into
the filter unit and connected to the LP C Laser
Printer (connection marked "20").

35

37

The filter unit is powered by the LP C
Cassette Printer.