20

Operating Manual V 1.0 – 06/2010

4.5

Installation site requirements

The installation site must be well-ventilated.
Never operate the instrument in rooms with explosion hazard.

4.

Instrument Setup

•

The printer needs an installation area of approx. 260 x 460 mm and the fil-
ter unit needs an additional installation area of approx. 470 x 490 mm.
Additional space is required for the PC containing the control software.

•

Stable laboratory bench and practically vibration-free floor.

•

Room temperature consistently between +5 °C and +40 °C.

•

Relative air humidity maximum 80 %, non-condensing.

•

Avoid vibrations, bright direct light and strong temperature fluctuations.

•

The AC socket used for the power supply must be close to the laser
printer and easily accessible.

•

The laser printer is intended for use in enclosed rooms only and may be
used at elevations up to 2000 m only.