2 cleaning the filter inlet tubes, Cleaning the filter inlet tubes – Leica Biosystems LPC Operating Manual User Manual

Page 27

Advertising
background image

27

Leica LP C

5.2

Cleaning the filter inlet tubes

The filter system of the Cassette Print-
er MUST be cleaned every 2 weeks to
maintain proper function.

The silver intake tube (29) is located inside the
main cover.

Cleaning the Intake Tube

The silver intake tube (29) and the exhaust shaft
(29a) collect dust if not cleaned every 2 weeks.

Use a gauze pad in a size of 5 x 5 cm to clean
the intake tube and the area around the tube.

Wipe this area with a DRY gauze pad!

Then, use a gauze pad moistened with 5%-8%
acetic acid (with distilled water) to clean the
intake tube and the area around the tube.

Fig. 24

29

5.

Cleaning and Maintenance

29a