Leica HI1220 Flattening Table

3. Instrument Components and Specifications

Nominal supply voltages:

Two factory-preset voltages (not user-adjustable):

100-120 VAC + 10 %

230-240 VAC + 10 %

Nominal frequency:

50/60 Hz

Nominal power:

350 VA

Main fuse:

2 melting fuses, 5x20 mm, UL-approved

For 100-120 V: 5x20 mm, 2x T 5A L250 VAC

For 230-240 V: 5x20 mm, 2x T 2.5A L250 VAC

Dimensions:

350x310x100 mm

Empty weight:

3.6 kg (w/o packaging)

Relative humidity:

20-80 %, non-condensing

IEC 1010 classification:

Protection class: 1

Pollution degree: 2

Overvoltage category: II

Operating temperature range:

+15 °C to +40 °C

Temperature control range:

Room temperature to 75 °C

Operating elevation:

maximum of 2000 m above sea level

IP protection class (IEC 60529):

IP20