Instrument Setup

4.1 Location conditions

• When installing the power cable, always be sure to route it so that it cannot contact the

heated surfaces of the instrument at any time.

• The power socket to which the instrument is to be connected to has to be located near the

instrument and easily accessible.

• The power supply must be at a distance no greater than the length of the power cable – an

extension cable must NOT be used.

• The substrate must be largely vibration-free and have sufficient load capacity and rigidity

for the weight of the instrument.

• Avoid impacts, bright direct light, and excessive temperature fluctuations.
• The instrument must be connected to a suitable power socket.Use only the provided power

cable, which is intended for the local power supply.

4.2 Unpacking and Setting Up the HI1210

Fig. 3

Fig. 2

Set the carton on a flat surface, cut through

the adhesive tape (

Fig. 2

) and open the

carton (

Fig. 3

).