Supplemental sheet for the ST5020 V1.9 Instructions for Use - 12/2012

Supplement to Chapter "7.2 Troubleshooting", page 78

7.2.1 Excessively high water level in the drain pan

Caution!

As long as the level sensor is covered with water, the water

flow cuvettes' (7-12) water valves are closed. The cleaning pro-

cess in the water flow cuvettes is interrupted. No fresh water

flows into the water flow cuvettes.

Supplement to Chapter "3.2 Instrument specifications", page 12 and

Chapter "9.2 Optional Accessories", page 85

3

Supplement ST5020 V1.3, RevD - 2/2013

Whenever the water level in the drain pan reaches the level sensor (

Fig. 2

, 1), an alarm mes-

sage is triggered and the water inflow into the water flow cuvettes (7-12) is stopped for safety

reasons. The water inflow is automatically turned on again whenever the water level no longer

covers the level sensor.

Sources of errors:
1. The water drain or the drain tube is blocked or clogged. The water

level in the drain pan has reached the level sensor; the water does

not drain.

2. The water drain or the drain tube is partially clogged or blocked. The

water level in the drain pan has reached the level sensor; the water

drains only slowly. The level sensor is dirty.

Troubleshooting
1. Read the error message, acknowledge the alarm and, immediately

afterward, press the "PAUSE" knob. The transport arm moves into a

secure position. The instrument is now in pause mode and the instru-

ment cover can be opened.

Subsequently, remove cuvettes 1 and 13 (

see label in Fig. 1

). Inspect

the water drain for clogs and the level sensor for dirt.

Fig. 1

To prevent damage to the instrument or the specimen, only

accessories authorized by Leica may be used.