7.2.1 Excessively high water level in the drain pan (continued)

Supplement to Chapter "7.2 Troubleshooting", page 78 (continued)

Supplemental sheet for the ST5020 V1.9 Instructions for Use - 12/2012

Caution!

If there are specimens in the instrument, these must be secured

and manually removed from the cuvettes and out of the instru-

ment by the user. Depending on the staining steps affected at

the time of the alarm, the user must secure the specimen so that

no damage to the specimen results (for example, by transferring

them into suitable, compatible reagents).

5

Supplement ST5020 V1.3, RevD - 2/2013

The programs already started that are to be assigned to the specimen

slide holders removed manually must be ended with the "CANCEL

BASKET" function (

refer to Chapter 5.8.4 - Quitting programs

).

Important information!

When ending the programs with the "CANCEL BASKET" func-

tion, a later proper resumption of the staining process for other

specimen slide holders is ensured.

If it is not possible to ensure a lasting and sufficient drainage of

water from the instrument by cleaning the water drain and the

drain tube, the instrument must be switched off and the respon-

sible Leica service department must be contacted.