Loadstar Sensors DI-1000U Calibration Quick Start Guide User Manual

Page 2

Advertising
background image

Loadstar Sensors, Inc. •

www.loadstarsensors.com

033-02129 Rev 2 2

7.

To calibrate using a known weight type the following commands in sequence. Commands
are not case sensitive. For example, to calibrate a 250 lb load cell using a known load of 100
lb, remove all load from load cell and type:

UNITS LB

LC 250

CAL 2

2PCAL 100
<
Apply Point 1 Load of 0 LB Press C when ready or Q to quit> C
<
Apply Point 2 Load of 100 LB Press C when ready or Q to quit> Apply 100 lb, then type C

TARE

WC (this will continuously output the load in lb. Change the load and the output
should change. Type <Enter> to stop the continous output.).

8.

To calibrate a load cell in

kg

using mVolt/V type the following commands in sequence in

Hyperterminal. Commands are not case sensitive. For example, to calibrate a 250 kg load
cell with a 3.02 mVolt/V rating:

UNITS KG

LC 250

CAL m

MVOLT 3.02 (This number will be available from the load cell
calibration/specification certificate.)

TARE (this zeros out the load)

WC (this will continuously output the load in kg. Change the load and ensure that
the output changes).

9.

To calibrate a load cell in

kg

using a known weight type the following commands in

sequence. Commands are not case sensitive. For example, to calibrate a 250 kg load cell
using a known load of 100 kg, remove all load from load cell and type:

UNITS KG

LC 250

CAL 2

2PCAL 100
<
Apply Point 1 Load of 0 KG Press C when ready or Q to quit> C
<
Apply Point 2 Load of 100 KG Press C when ready or Q to quit> Apply 100 kg, then type C

TARE

10.

WC (this will continuously output the load in kg. Change the load and the output should
change. Type <Enter> to stop the continuous output.).