Installation Procedure

Model 685.53 Grip Control Module

Installation

13

Do not clamp the grips without a specimen installed.

Some grips may be damaged by closing the grip without a specimen
inserted or with a specimen only partially inserted.

See the appropriate grip product information manual for detailed specimen
mounting instructions.

10. Check the grip operation.

Change the grip controls between grip and release a few times and check for
leaks at all connections (correct as needed).

11. Secure the control panel.

Lift the control panel back up to the base and secure with the screws
removed in Step

4

. Do not overtighten.

12. Replace the baseplate cover.