MTS Model 685.53 Grip Control Module User Manual
Page 17

Controls
Model 685.53 Grip Control Module
Operation
17
Pressure
Adjusts the amount of hydraulic pressure applied to the grips for gripping the
specimen.
This is a multiturn knob. Pressure increases as the knob is turned clockwise
and decreases as the knob is turned counterclockwise.
To determine the correct operating pressure, see
Rate
Adjusts how fast the grips clamp and unclamp. This control requires the use of
a hex-key wrench for adjustment.
Adjust the Rate control clockwise to slow the clamping of the specimen.
Otherwise, adjust the Rate control counterclockwise to increase the clamping
speed.
To adjust the clamping rate, see
Pressure Gage
Shows the current hydraulic pressure setting for grip operation.
The gage includes scales for both MPa and psi units.
Use the gage while adjusting the Pressure control to achieve the desired
clamping pressure. See the grip product information manual for gripping
pressures.
Control
Description
Pressure
Rate
10
1
5
6
7
9
4
2
8
3