7.

Setup Configuration

To change or view the setup configuration, the Pump must be stopped. Press and hold the ‘Diameter’/‘Setup’
key until the parameter, ‘PF’ is displayed. After 2 seconds, or when any non-arrow key is pressed, the next
parameter will be displayed. (See sec. 5.10, ‘Setup’ Key). Pressing an arrow key under a value will change the
parameter. The following will be displayed:

n

Power Failure Mode. Where ‘n’ is the current setting: ‘0’ = Disabled, ‘1’ = Enabled.

When enabled, if the Pump was pumping when power to the pump was disrupted, the Pump will automatically
start pumping when power is reconnected to the pump. The accumulated volume will be reset to 0. Pressing
any key on the keypad while powering up the pump will stop the Pump from starting in Power Failure Mode.

n

Low Noise Mode, Where ‘n’ is the current setting: ‘0’ = Disabled, ‘1’ = Enabled.

A side effect of the NE-300’s high precision micro-stepped motor driver is a high frequency resonance sound
at very low pumping speeds. This mode minimizes this sound by reducing the micro-stepping, increasing
pulsations.

8.

Rear of Pump

NEW ERA PUMP SYSTEMS, Inc.

WWW.SYRINGEPUMP.COM

(631) 249-1392

Farmingdale, NY [email protected]

NE-300

Figure 2: Rear of Pump

9.

Appendix

9.1

Accessories

9.1.1

Syringe Heater

See www.SyringeHeater.com for details.

Flexible heating pads that wraps around the syringe. Thermo-Kinetic Heat Clamping controller will heat a
syringe to a set temperature up to 100 C.

9.1.2

Firmware and Hardware Upgrade

Contact your dealer about upgrading your pump with other features from the NE-1000 Programmable Syringe
Pump Series. See www.SyringePump.com