Setup, Loading syringes, Plunger flange plunger barrel barrel flange – New Era NE-300 Just Infusion Syringe User Manual

Page 6

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-300 Syringe Pump

Copyright: New Era Pump Systems Inc.

3

08/27/14

To facilitate and enhance your understanding of the NE-300’s operation, please take the time to familiarize
yourself with the basic concepts below:

Parts of the Pump

1) Pusher Block
2) End Plate
3) Power On/Off Switch
4) Drive-Nut Button
5) Drive-Screw

6) Syringe Holder Block
7) ‘V’ Slot (on Syringe Holder Block)
8) Syringe Clamp
9) Keypad / User Interface
10) Guide Rod (2 guide rods)

Parts of a Syringe

Terminology

Momentary Press:

A quick press, less than 1 second, and then release of a key on the keypad.

Display Blink:

A momentary blanking of the LCD display. This indicates that the new data
entered by the user is valid and has taken effect.

3.

Setup

Place the pump on a stable surface.

Plug the round connector end of the supplied power supply adapter into the power plug located on the
lower right of the pump's rear. See section 8, Rear of Pump, for a diagram of the rear of the pump. Plug
the other end of the power supply adapter into an appropriate electrical outlet. The pump will be powered
when the bottom of the power switch, located on the upper right of the rear of the pump, labeled ‘1’, is
pressed. The red indicator on the switch is visible when the power switch is in the ‘on’ position. After
power is applied to the pump, the pump’s display will flash.

Next, the Pumping Parameters can be entered. Before the pump can be started, the pump needs the
measurement of the inside diameter, in millimeters (mm), of the syringe that will be loaded. The syringe
diameter is entered using the keypad on the front panel of the pump.

Finally, the syringe can be loaded and the pump started.

4.

Loading Syringes

The syringe is loaded by securing the barrel and the pusher flange as follows:

1: Press in fully the white drive-nut button on the pusher block, releasing the block. Taking care not to

drag the drive-nut on the drive-screw, slide the block away from the syringe holder, providing sufficient
space for the loaded syringe. Then release the white button.

2: Lift the syringe clamp above the syringe holder block. Turn it 1/4 turn and then lower it onto the

syringe holder block. The syringe clamp should be out of the ‘V’ slot.

3: Load the syringe with the barrel over the syringe holder and the syringe plunger towards the middle of

the pump. Place the barrel on the syringe holder, in the ‘V’ slot, with the barrel flange to the left of the
syringe holder block.

4: Lift the syringe clamp to slightly above the height of the syringe barrel and then turn the syringe clamp 1/4

turn back to its original position and lower it onto the syringe barrel.

5: Press the white drive-nut button to slide the pusher block against the syringe plunger. Release the white

drive-nut button.

To unload the syringe, reverse the instructions for syringe loading.

Plunger
Flange

Plunger

Barrel

Barrel Flange