Troubleshooting and maintenance, Specifications mechanical & electrical, Operational – New Era NE-300 Just Infusion Syringe User Manual

Page 13: Custom applications

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-300 Syringe Pump

Publication #1200-01

10

08/27/14

9.2

Troubleshooting and Maintenance

Maintenance: Periodically, a small amount of all-purpose oil should be applied to the guide rods.

The mechanism should be kept clean to prevent impeded operation.

No other special maintenance or calibrations are needed

Pusher block makes a snap or click sound when the pump is started: This is a normal condition. When
the pusher block is manually moved, the drive-nut may not have been fully engaged on the drive screw. The
sound heard is the drive-nut engaging on the drive screw.

9.3

Specifications

Mechanical & Electrical

Syringe sizes:

Up to 60 mL. 140 mL partially filled.

Number of syringes:

1

Motor type:

Step motor

Motor steps per revolution:

400

Micro-stepping:

1/8 to 1/2 depending on motor speed

Advance per step:

0.2126 µm to 0.8504 µm depending on motor speed

Motor to drive screw ratio:

15/28

Drive screw pitch:

20 revolutions/”

Power connector:

2.1 mm, center positive, DC

Voltage at power connector:

12V DC at full load

Amperage:

750 mA at full load

Power supply type:

External wall adapter, country and power source specific

Power supply output rating:

12V DC @ 800 mA (or greater if regulated power supply)

Dimensions:

8 3/4” x 5 3/4” x 4 1/2” (L x W x H)
(22.86 cm x 14.605 cm x 11.43 cm)

Weight:

3.6 lbs. (1.63 kg)

Operational

Accuracy

Within +/- 1% over length of syringe,
exclusive of syringe variations.

Reproducibility

+/- 0.1%


Maximum speed:

3.7742 cm/min

Minimum speed:

0.004205 cm/hr

Maximum pumping rate:

1257 mL/hr with a B-D 60 mL syringe

Minimum pumping rate:

0.73 µL /hr with a B-D 1 mL syringe

Maximum force:

35 lbs. at minimum speed, 18 lbs. at maximum speed

Syringe inside diameter range:

0.100 to 50.00 mm

9.4

Custom Applications

For specialized and OEM applications, contact your dealer
or New Era Pump Systems Inc. Custom modifications can
be made to the mechanics or the firmware.

Advertising