Default Program Selection
Use one pump as the Master control pump, and the other as the Secondary slave pump. After
selecting a default program, the default parameters can be modified. Do not load program on
second pump. Set the syringe diameter on both pumps.
Basic Pump Setup: Select Default Dual Pumps Program
On the Master Pump ONLY: Turn power off.
Press and hold the “Program Function” key (Volume key).
Turn power on, and then release the “Program Function” key.
Press any up-arrow key to select a default program.
Default program will be loaded and the communications mode will be set.
Selectable Default Programs
Description
Displayed as
‘X Version Only
Reciprocating Pumps / Auto Refill
rECP
Next Generation Continuous Infusion
COnt
Gradient Inverse Linear Constant Rate
GrAd
Plumbing for a Continuous Infusion System (P-DKIT)
Major Parts of the System
Quantity
P-DKIT
Description
2
NE-1000 Series Syringe Pump
1
CBL-DUAL-3 control cable
2
Syringes (60 mL)
2
Dual Check Valve
2
Y Connector
6
Sections of Tubing