Alternating pumps mode – New Era CBL-DUAL-3 RS-232 Dual pumps communications cable (3) User Manual

Page 4

Advertising
background image

4

Alternating Pumps Mode

Creates a semi-automated continuous infusion system using pre-loaded syringes. The infusing
pump will automatically start the alternate pump when the set volume is reached. The empty
syringe can then be manually replaced during the infusion of the alternate pump.

Alternate pump starts when first pump stops.

‘X Version Firmware Only (Enhanced features)

• Delay Start of second pump.

Set the TTL setup setting: RUN.1

Delays start of alternate pump while executing a Pause Function. Alternate pump will start when
the Pumping Program has stopped.

• Overlap pumping of the second pump.

Pumping Program Function: OUT.0

Immediately sends start to alternate pump when OUT.0 function is executed.

Allows the creation of an overlap between the infusing and the alternate pump to allow the
alternate pump to prime the syringe before the infusing pump stops.

If Phase 1 is set to function OUT.0, the alternate pump will immediately start when the Master
pump starts.

Overrides sending a start command when the infusing pump stops.

• Stop Alternating Mode

Set TTL Trigger Mode to “Off”. When trigger is set to “Off”, start command will not be sent to
alternate pump.

Within a Pumping Program, the trigger mode can be changed with the Trigger function. This
allows a Pumping Program to limit the number of times that the alternate pump will execute its
Pumping Program.