Default Continuous Pumping Program

To load the default program: On the Master pump, while turning on power:

Press and hold the “Volume”/”Program Function” key until [Cont] is displayed. Press the
“Volume” button again to enter the selection.

Default Continuous Mode Pumping Parameters

Phase #

Function

Rate

Volume

1

RATE

Infusion

100.0 mL/hr Syringe

10.00 mL

2

RATE

Refill

500.0 mL/hr Overlap 0.000 mL

3

RATE

Prime

200.0 mL/hr Prime

0.100 mL

4

RATE

Prefilled mode 0.000 mL/hr (not used) 0.000 mL

Address Mode set to Continuous: [Cont]