Measure with mam sensor – RaySafe X2 Manual User Manual
Page 26
Advertising

MEASURE WITH MAM SENSOR
The MAM sensor is ready to measure dose and HVL for all
anode/filter combinations without changing any settings.
Place the connected sensor centered in the field, with the
crosshair 6 cm (4 cm in the US) from the examination table
front edge (A). The angle in the horizontal plane has
neglectable impact on the measurement result.
For kVp measurements, select your anode/filter combination
in the sensor settings, or by swiping right from the home
screen to access the quick settings.
Note!
When measuring kV on Mo/Rh, use the included 2 mm Al filter. Follow the instructions printed on
the filter.
Expose.
Read the result.
RaySafe X2 User Manual