MO/RH KV WITH MAM SENSOR

Place the sensor centered and 6 cm from the chest wall.

Place the 2 mm Al filter on the sensor. Be sure to cover the

active sensor area.
Recommended generator setting ≥ 100 mAs.
Select Mo/Rh 2 mm Al in the sensor settings.