W/AL SCANNING WITH MAM SENSOR

Mount the X2 MAM sensor in the holder.

Place the holder centered on the table, tight to the chest

wall.

Settings:
X2:

Generic (W/Al if kVp is desired), Stop delay: 2 s

Philips MicroDose:

8 mAs

Fischer SenoScan:

70 mA

Adani:

fastest scan speed

Always use the compression paddle, positioned as high as possible. Do not use the AEC functionality.