Replacing the breaker jaws
Fig. 10
The stationary breaker arm S can be removed to provide
sufficient space for dismantling the machine. Fig. 10
*
Unscrew the machine screws F.
*
Swing back the fill hopper B.
*
Grasp the breaker arm S at the grip and pull out the pin
Q.
*
Lift out the breaker arm S.
*
Unscrew the machine screws I.
*
Remove the clamping shims H.
*
Replace the breaker jaws J.
*
Assemble in reverse order.
*
Set the gap width as described in this instruction manual.
!
The fill hopper may be swung back (by unscrewing the two
machine screws) only by trained personnel and only for
cleaning purposes. Operating the BB 200 without these
machine screws in place and tightened down properly is
absolutely prohibited.