Operating the machine

20

Fig. 4: Level Display

Determination of the level in the vessel

Read the Dewar flask level on the scale in the level display.

Alternatively, you can use a dipstick to determine the level and compare it
with the level curve (see picture of Apollo® 50 level curve).

The black area on the level display scale marks 25% of the remaining contents.

Fig. 5: Level curve Apollo ® 50

Pos : 6.16 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Füll en des Behälters @ 2 \mod_1267774868556_9.doc @ 18870 @ @ 1

4.6 Filling the vessel

Pos : 6.17 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1603 Modul Dewar 50l Befüllen @ 2 \mod_1267698001736_9.doc @ 18834 @ @ 1

Close all valves. (V1, V2, V3)

Take the withdrawal hose (ZF) off the connection (SA).

Attach the filling line to the connection (SA).

Open the waste gas / overflow valve (V3).

Open the withdrawal valve (V1).

Fill the Dewar flask until liquid nitrogen comes out of the waste gas /
overflow valve (V3).

Close the withdrawal valve (V1).