Operating the machine, 7 pressure build-up, 8 withdrawal of liquid nitrogen – Retsch CryoMill Autofill 50l User Manual

Page 21: 9 pressure relief, 10 transport, Caution

Advertising
background image

Operating the machine

21

Close the waste gas / overflow valve (V3).

Pos : 6.18 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Druc kaufbau @ 1\mod_1240400926229_9.doc @ 8240 @ @ 1

4.7 Pressure build-up

Pos : 6.19 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1604 Modul Dewar 50l Druc kaufbau @ 2\mod_1267701477789_9.doc @ 18841 @ @ 1

Close the withdrawal valve (V1) and the waste gas / overflow valve (V3).

Open the pressure build-up valve (V2).

Watch the pressure indicator (DM).

Close the pressure build-up valve (V2) as soon as the needle reaches the
red marking or nitrogen comes out of the pressure relief valve (PSV).

Fig. 6: Pressure gauge

Pos : 6.20 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Entnahme von Flüs sigstic kstoff @ 2\mod_1267774957314_9.doc @ 18877 @ @ 1

4.8 Withdrawal of liquid nitrogen

Pos : 6.21 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1605 Modul Dewar 50l Entnahme Flüssigstic kstoff @ 2\mod_1267709072517_9.doc @ 18855 @ @ 1

Attach the liquid nitrogen line to the CryoMill and to the connection (SA).

Open the filling and withdrawal valve (V1).

After taking the liquid, close the filling and withdrawal valve (V1) again.

Pos : 6.22 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Druc kentl astung @ 2\mod_1267777113490_9.doc @ 18891 @ @ 1

4.9 Pressure relief

Pos : 6.23 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1606 Modul Dewar 50l Druc kentlastung @ 2\mod_1267709470983_9.doc @ 18862 @ @ 1

Close the pressure build-up valve (V2).

Open the waste gas valve (V3).

Pos : 6.24 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Trans port @ 0\mod_1226495164391_9.doc @ 3398 @ 2 @ 1

4.10 Transport

Pos : 6.25 /00004 Warnhinweis e/V0019 VOR SICHT Überdr uc k i m Behälter - Gefärdung durc h LN 2 @ 2\mod_1267775099915_9.doc @ 18884 @ @ 1

CAUTION

Risks of liquid or gaseous nitrogen
Positive pressure in the vessel

The introduction of energy during transport can cause a sudden surge in
pressure. Nitrogen is blown off then and this can cause a lack of oxygen,
cryogenic burns, oxygen enrichment and associated risks.

Make sure the vessel is depressurised before you transport it,

Remove the siphon and put in the transport plugs.

Transport the vessel only in an upright position.

Avoid impacts and vibrations.

Secure the vessel against falling down, rolling away and damage.

Pos : 6.26 /00005 Ü berschriften/1. Ü berschriften/1 Prüfanweis ungen @ 1 \mod_1245230896441_9.doc @ 11114 @ @ 1

Pos : 6.27 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1612 Modul Dewar 50l Verboten ist die An wendung folgender Mittel und @ 2\mod_1268231540473_9.doc @ 18957 @ @ 1

It is impermissible to use the following agents and tools:

pliers

striking tools

DM

Advertising