Retsch AS 300 User Manual

Page 13

Advertising
background image

07/02/2012

Retsch GmbH

13

Intermittent – continuous operation

7

8

9

10

11

When the AS300 control is switched on, intermittent operation is ini-
tially disabled.

Enabling intermittent mode of 10-99 sec. :

Press key 10
The LED above key 10 illuminates
Display 8 shows 10 sec.
Pressing key 9 raises the interval period up to 99 sec.

If 99 sec is overshot, start again with 10 sec.

Key 7 reduces the interval period down to 10 sec.

If 10 sec. is undershot, start again with 99 sec.

Disabling intermittent mode :

Press key 11
The LED above key 11 illuminates

Display 8 is extinguished

Memory – saving and calling-up sieve parameters

2

5

8

10 11

13

The memory function, thus saving or calling-up preselected sieve pa-
rameters, is only available in standby mode. You are able to change
and overwrite program places P1 to P9 at any time.

After the AS300 control has been turned on, display 2 shows "on".
The next program place P1 to P9 is reached by pressing the PROG
key. on reappears in the display after program place P9.

on = the sieve parameters 5/8/10-11 and 13 can be freely
set here.

or

P1 to P9 = sieve parameters can be stored and called-up here

Start = sieving process is started with the sieve parameters
stored in program place P1 to P9.

Occupying storage space

Press the PROG key until the required storage place P1-P9
has been reached.

Press the SET key, all displays flash

Set sieve parameter 5/8/10-11/13

The programming mode can be aborted by pressing the PROG button
– the values are not stored.

Press SET button – the values are stored

Flashing of the display stops, the adjusting interlock is activated, the
sieving parameters are stored.

If the AS300 control is actuated with a PC and the "EasySieve

®

"

sieve software, display 2 shows "ES". No manual changes to the sieve
parameters are possible in this mode.