Retsch AS 300 User Manual

Page 14

Advertising
background image

07/02/2012

Retsch GmbH

14

Vibration height in "mm"

A value twice the amplitude is shown as the measured value in dis-
play 5 marked amplitude. We refer to this measured value
throughout these instructions as the vibration height.
When the AS300 control is switched on, the vibration height is
preset to 1.00 mm.

Setting the vibration height from 0.20 – 2.20mm :

4

5

6

Reduce :

Key 4 reduces the vibration height down to 0.20 mm
Hold down key 4, after 5 sec. the rapid adjustment function will

commence

If 0.20 mm is undershot, 2.20 mm appears again

Increase :

Key 6 increase the vibration height up to 2.20 mm
Hold down 6, after 5 sec. the rapid adjustment function will com-

mence

If 2.20 mm is overshot, 0.20mm appears again

Sieve plate acceleration in "g"

When the AS300 control is switched on, the vibration height is
preset to 1.00 mm and can be switched to sieve plate accelera-
tion mode by simultaneously pressing and holding down

keys 4+6 for 2s

After switchover, display 5 still does not show any values, this
only happens after the START has been pressed.
The display 5 marked as amplitude mm will show a multiple of
the Earth's gravity "g" as the measured value.
( 1g = 9,81m/s

2

)

Setting the sieve plate acceleration in "g"

from 1 – approx.17g

Naturally, your AS300 control can only achieve the sieve

plate acceleration at which, in dependence on the inherent fre-

quency, an amplitude of 2.2 mm is not overshot.

4

5

6

Reduce :

Key 4 reduces the sieve plate acceleration down to 1g
Hold down key 4, after 5 sec. the rapid adjustment function

will commence

Increase :

Key 6 increase the sieve plate acceleration up to approx. 17g
Hold down 6, after 5 sec. the rapid adjustment function will

commence

Signal tone for the end of sieving

11

15 16

If the sieving process runs to the end without interruption, the end
is then signalled by an acoustic tone sounding 5 times.

Disabling the signal tone :

Press keys 11 and 16 simultaneously

confirmed by a signal.

Enabling the signal tone :

Press keys 11 and 15 simultaneously

confirmed by a signal.