Model GP-01 Operator’s Manual

Maintenance • 25

Maintenance

This section describes troubleshooting procedures for the Model GP-01. It
also describes how to change or charge the Model GP-01’s batteries as
well as how to replace the unit’s gas sensor and filter disks.

WARNING: RKI Instruments, Inc. recommends that service,

calibration, and repair of RKI instruments be performed
by personnel properly trained for this work. Replacing
sensors and other parts with original equipment does not
affect the intrinsic safety of the instrument.

Troubleshooting

The troubleshooting table describes error messages, symptoms, probable
causes, and recommended actions for problems you may encounter with
the Model GP-01.

Table 6: Troubleshooting the Model GP-01

Symptoms

Probable

Causes

Recommended Action

• The LCD is

blank.

• The unit may

have been
turned off.

• The alkaline

batteries may
need to be
replaced or the
Ni-cad
batteries
recharged.

1. To turn on the unit, press

and hold the POWER/
MODE button.

2. If the unit does not turn on,

replace the alkaline
batteries or recharge the
Ni-cad batteries.

3. If the difficulties continue,

contact RKI Instruments,
Inc. for further instruction.

• The LCD shows

abnormally high
readings but
other gas
detection
instruments do
not.

• The unit may

need to be
recalibrated.

• The sensor

may need
replacement.

1. Recalibrate the unit.
2. Replace the sensor and

calibrate the unit.

3. If the difficulties continue,

contact RKI Instruments for
further instruction.