Replacing the combustible sensor – RKI Instruments GP-01 with self resetting alarms User Manual

Page 29

Advertising
background image

Model GP-01 Operator’s Manual

Maintenance • 29

3. When a full charge has been reached, approximately 40 minutes, the

red LED on the charging station will turn off. Remove the battery pack
from the charging station and unplug the charging station’s power cord
from the AC outlet.

Replacing the Combustible Sensor

WARNING: Replace the sensor in a non-hazardous environment.

1. Verify that the Model GP-01 is off.

2. Remove the sensor end protective rubber boot.

3. With a small Phillips screwdriver, carefully unscrew the two screws that

hold the sensor retainer in place.

Red Charging
LED

Battery Pack

Figure 9: Putting Battery Pack Into Charging Station

Sensor Retainer

LEL Sensor

POWER

MODE

AIR

GP-01

%LEL

Figure 10: Removing the Sensor Retainer