Specifications – RKI Instruments GP-01 with self resetting alarms User Manual

Page 6

Advertising
background image

6 • Specifications

Model GP-01 Operator’s Manual

Specifications

Table 1: Model GP-01 Specifications

Target Gas

Combustible Gas, Methane (CH

4

) Calibration Standard

Detection Range

0-100% LEL (1% LEL Increments)

Detection Principle

Catalytic Combustion Method

Sampling Type

Diffusion

Response Time

T90 in 30 seconds

Indication

7-Segment Digital LCD

Gas Alarms
(Factory Settings)

•

Low Alarm: 10% LEL

•

High Alarm: 50% LEL

•

Over Range: 100% LEL

Other Alarms

Sensor Failure, Low Battery, Dead Battery, System Failure, and
Calibration Failure

Operating Temperature &
Humidity

- 20°C — 50°C, Below 90% RH (Non-condensing)

Power Supply

•

Two AAA size Alkaline Batteries Standard, Duracell MN2400 or
PC2400, Eveready Energizer E92 or EN92

•

2.4 Volt Ni-cad Battery Pack 49-1603RK Optional

Continuous Operating Time

•

Alkaline Batteries: 16 Hours Minimum at 25°C, No Alarms or
Backlighting

•

Ni-cad Battery Pack: 8 Hours Minimum at 25°C When Fully
Charged, No Alarms or Backlighting

Features

•

Manual Backlight on LCD (Automatic Lighting at Alarm)

•

Peak Hold

•

Direct Charging (Ni-cad Battery)

Standard Accessories

•

Wrist Strap

•

Rubber protective boots

Optional Accessories

•

Ni-cad Battery Pack

•

Battery Charger for Ni-cad Battery Pack

•

Calibration Adapter

•

Calibration Kit

•

Alligator Clip

•

Belt Clip

Safety/Regulatory

C US

186718

CSA certified, “C/US”, as Intrinsically Safe. Exia. Class I, Div. 1, Groups

A, B, C, & D. Temperature Code T3C.

Dimensions and Weight

1.4” (35mm) W x 4.1” (104mm) H x 0.8” (20mm) D; 2.8 oz. (80 g)

Advertising