8 • Internal Description

RI-215A

Operator’s Manual

On the bottom of the front housing is a series of slots and a calibration port. The slots
allow the target gas sample to enter the IR gas sensor in the RI-215A. The calibration port
is covered with a rubber cap. The calibration port accepts

1

/

4

-inch ID (inside diameter)

flexible tubing. A series of switches are accessible to the right of the calibration port, as
shown in Figure 1 on page 7. These switches include:

Mode (M)

Up/Alarm (

)

Down (

)

Set/Calibration (S)

The base plate has a wire entry slot for the cable assembly. The cable assembly connects
directly to the printed circuit board inside the case through the cable entry slot. The cable
assembly is used to connect to an external power source. You also use it to make
connections to an external alarm and to a monitoring or recording device. The base plate
has two mounting holes for convenient mounting on vertical surfaces at the monitoring
site.

Cable Assembly

The cable assembly connects to a six-pin receptacle on the printed circuit board through a
slot on the rear of the case. It allows you to make external wiring connections to the RI-
215A.

Internal Description

Printed Circuit Board

On the printed circuit board is located the IR gas sensor, LCD, and fuse. Connected to the
printed circuit board is the cable assembly.

The printed circuit board converts the output of the IR gas sensor to a 4-to-20 mA signal
(that is proportional to the detection range of the RI-215A). The printed circuit board also
displays a digital reading on the LCD.

The printed circuit board has four switches: Mode (M), Up/Alarm (

), Down (

), and

Set/Calibration (S). These switches are located at the bottom of the printed circuit board
beneath the gas sensor. Table 2 lists each switch and its function.

Table 2: Model RI-215A Program Switches

Switch

Function

Mode (M)

Enters Maintenance Mode

Up/Alarm (

)

Increases the setting on the LCD

Displays the alarm point

Down (

)

Decreases the setting on the LCD

Set/Calibration (S)

Prepares a setting to be changed

Enters a new setting