6 • Operation

Model GX-82 Instruction Manual

Normal Operation

The GX-82 continuously monitors the atmosphere and displays the LEL, O

2

, and H

2

S or

CO concentrations present. If the GX-82 is taken into a low-light environment, the display
backlight automatically turns on.

Displaying battery voltage

Press the BATT/DOSE button. A single tone sounds and the battery voltage displays.
New alkaline batteries will register approximately 3.0V; freshly recharged Ni-Cd batteries
will register approximately 2.7V. The minimum voltage required for the GX-82 is 2.3V.

NOTE:

The GX-82 automatically checks battery voltage during start-up. If the
measurement is below 2.3V, the GX-82 will not operate.

During normal operation, when the battery voltage drops to 2.3V, the letter

B

flashes on the display. When the voltage reaches 2.1V, all four LEDs turn on, a

continuous alarm tone sounds, and

LOW BATT

displays.

WARNING: The GX-82 does not monitor the atmosphere during the low battery alarm.

When the

B

indicator flashes on the display, take the GX-82 to a non-

hazardous location to change the batteries. (See Maintenance-Batteries.)

Displaying toxic gas (H

2

S and CO) exposure

Press the BATT/DOSE button until a double tone sounds. The GX-82 displays

D= XXX

ppm *XXXM

. This display indicates the H

2

S or CO average exposure (

D= XXX ppm

), and

duration of the measurement in minutes (

*XXXM

). The GX-82 returns to normal operation

after five seconds. The duration of the measurement is the time elapsed since the last time
the GX-82 was turned on.

Displaying peak hold

Press the PEAK/HOLD button. A single tone sounds and the letter

P

and readings for all

three gases display. While

P

is displayed, the GX-82 updates and displays the peak

reading (lowest for O

2

) for each channel since the time you pressed the PEAK/HOLD

button.

To return to normal operation, press the PEAK/HOLD button again. The GX-82 continues
to update and display peak readings until you return to normal operation or turn the
GX-82 off.

NOTE:

The BATT (battery voltage) and DOSE (toxic exposure) functions do not work

while the GX-82 is in the PEAK hold condition.

Auto zero/span

To automatically adjust the zero point for combustibles, H

2

S or CO, and the span for

oxygen, take the GX-82 to a known fresh-air environment. With the GX-82 in normal
operation, press and hold the PEAK/HOLD button until a single tone sounds, then also
press and hold the BATT/DOSE button. The GX-82 displays

AUTO ZERO/SPAN

, which

indicates the GX-82 is adjusting the detection circuits based upon the fresh air reading.
When

AUTO ZERO/SPAN

displays, release the PEAK/HOLD and BATT/DOSE buttons.