Model GX-82 Instruction Manual

Safety Self-Check

The GX-82 continuously monitors itself for proper operation. If a malfunction occurs, a
single steady “trouble” tone sounds, and one of the following messages display.

Emergency Operation

The GX-82 can operate temporarily with up to three failed sensors. If a sensor(s) fails or is
absent at start-up, the GX-82 sounds a seven-tone alert, displays

FAIL COMB O2 CO

or

H2S

(any combination of two) then adjusts to operate on the remaining sensor(s). The GX-82
displays

XX

for the defective sensor(s) and normal readings for the active sensors. The

GX-82 performs normal monitoring functions for the active sensors for 13 minutes, when
it performs a self-diagnosis, then shows a malfunction alarm. To continue using the
GX-82, turn off the power and restart.

If a sensor fails during operation, the GX-82 shows a malfunction alarm. To continue
monitoring with the remaining active sensors, turn off the power and restart as described
above.

Table 4: GX-82 Error Messages

Display

Description

FAIL (COMB)

Indicates sensor failure (the failed sensor is
displayed).

00 OVER 00.0

Indicates O

2

concentration in excess of 40.0% or

an O

2

sensor failure.

WARNING: High oxygen concentration

increases the risk of fire hazard.

LOW BATT

Indicates the battery voltage is too low to run the
GX-82. See “Displaying battery voltage” on
page 6.