Model GX-82 Instruction Manual

Components and Controls • 3

CO and H

2

S sensor

The CO and H

2

S sensors are physically identical, except for their external labels. The

sensors are housed in a cylindrical plastic shell, with a bayonet-pin base for easy
replacement. The perforations at the end of the shell allow atmosphere to diffuse into the
sensor. The CO sensor also has a slip-on charcoal filter that covers the diffusion end. The
filter eliminates interference from H

2

S.

CO and H

2

S sensor - principle of operation

The CO and H

2

S sensors are electrochemical cells, consisting of two precious metal

electrodes in an acid electrolyte. A gas permeable membrane covers the cell and allows
gas in the atmosphere to diffuse into the electrolyte at a rate proportional to its partial
pressure. The gas reacts in the cell, producing a current proportional to the concentration
of gas. The current is amplified by the circuitry of the GX-82 and converted to a
measurement of gas concentration. To maintain sensitivity and stability, the CO and H

2

S

sensors are exposed to a continuous bias current, even when the power is off.

Control Panel

The control panel is at the top of the GX-82 for easy access when you carry the GX-82. The
control panel contains the display, alarm lights, and buttons that control the many
functions of the GX-82. The buttons are touch-pads to reduce the possibility of accidental
activation or damage.

Table 2: GX-82 Button Functions

Button

Function

POWER ON/OFF

The POWER ON/OFF button controls power to the GX-82.

(Bias power for the electrochemical sensors is not affected by

the POWER button.)

BATT/DOSE

The BATT/DOSE button displays the battery voltage and

the average level of H

2

S and CO exposure since the last time

the GX-82 was started.

PEAK HOLD

The PEAK/HOLD button displays the highest reading for

exposure to LEL and H

2

S or CO, and the lowest reading for

O

2

exposure. It is also used (with the BATT/DOSE button)

to activate the GX-82’s Auto Zero function.
(All measurements are tracked since the last time the GX-82

was turned on.)

Alarm Lights

Red light-emitting diodes (LEDs) labeled COMB, OXY, H

2

S,

or CO flash when the alarm for that gas is activated. The

alarm light automatically resets when the conditions fall

below (or rise above for O

2

) the alarm levels.