RKI Instruments 35-3000RK-LEL User Manual

Page 17

Advertising
background image

35-3000RK-LEL Combustilbe Gas Sample-Draw Detector • 17

NOTE: Make sure the bottom ferrule is laying flat in the nut.

6. Insert the cone-shaped front ferrule on top of the bottom ferrule. Insert the ferrule so

the smaller end of the cone is facing up.

7. Screw the nut onto the fitting by hand just until it stops turning.

8. Insert the sample tubing into the fitting until it bottoms out.

9. Firmly tighten the nut so the ferrules crimp onto the sample tubing and make a seal.

Adjusting the Low Flow Setting

NOTE: Adjusting the low flow setting will cause a low flow alarm at the sample-draw

detector and a fail alarm at the controller. Be sure to put the controller into its
calibration program or disable external alarms before performing this test

The factory-set low flow setting is 0.6 SCFH (±0.2). To adjust the low flow setting:

1. Use the flow adjust potentiometer (VR1) to set the flow to 0.6 SCFH.

If the sample-draw detector goes into low flow alarm before you can adjust the flow
down to 0.6 SCFH, adjust the low flow potentiometer 1/4 turn clockwise, then
attempt to set the flow again. Repeat this step until you are able to adjust the flow to
0.6 SCFH.

NOTE: The low flow potentiometer is accessible through a circular cutout in the main

circuit board. The cutout is labeled PS1.

2. Slowly turn the low flow potentiometer counterclockwise just until the sample-draw

detector goes into low flow alarm.

3. Increase the flow using VR1 until the unit is out of low flow alarm.

4. Decrease the flow very slowly and verify that the low flow alarm is 0.6 SCFH (±0.2).

If the low flow alarm is set too low, turn the low flow potentiometer slightly
clockwise. Repeat steps 3 and 4 if necessary.

5. Use the flow adjust potentiometer (VR1) to set the flow to 1.5 SCFH.

6. Make sure the sample-draw detector’s Fail light is off.

Advertising