Calibration frequency, Calibration, lel detector – RKI Instruments 35-3010RK-03 User Manual

Page 22

Advertising
background image

18 • 35-3010RK-03 Sample-Draw Detector

NOTE:

The low flow potentiometer is accessible through a circular cutout in the main
circuit board. The cutout is labeled PS1.

3.

Verify that the low flow alarm is 0.6 SCFH (±0.2). Repeat steps 3 and 4 if necessary.

4.

Use the flow adjust potentiometer (VR1) to set the flow to 1.2 SCFH.

5.

Make sure the sample draw detector’s Fail LED is off.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 months is adequate for most sample draw detector
applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 3 months for the sample
draw detector.

If an application is not very demanding, for example detection in a clean, temperature
controlled environment, and calibration adjustments are minimal at calibration, then a
calibration frequency of every 6 months is adequate for the sample draw detector.

If the application is very demanding, for example if the environment is not well
controlled, then more frequent calibration than every 3 months may be necessary for the
sample draw detector.

Calibration, LEL Detector

This section describes how to calibrate the LEL sensor in the sample draw detector. It
includes procedures to set the zero reading, set the response reading, and return to normal
operation.

NOTE:

This procedure describes calibration using a demand flow regulator, a zero air
calibration cylinder, and a 3-gas calibration cylinder.

Setting the Zero Reading

NOTE:

If you can verify a fresh air environment, it is not necessary to use the zero air
calibration cylinder to set the zero reading.

1.

Open the housing door.

2.

Screw the regulator into the zero air calibration cylinder.

3.

Connect the calibration kit sample tubing to the regulator.

4.

The zero reading for the LEL channel is set at the controller. See the controller
operator’s manual for instructions to set the zero reading.

5.

When the instructions call for applying gas to the detector, connect the sample tubing
from the regulator to the inlet line at or near the INLET fitting.

6.

Allow the gas to flow for 1 minute.

7.

Set the zero reading according to the controller operator’s manual.

8.

Disconnect the sample tubing from the inlet line.